ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 61

Advertising
background image

E Series Service Manual

9650-1450-01 Rev. D

2-9

ECG LEAD OFF

One or more ECG leads are not properly connected when
leads are selected as input.

Check cable and patient connection. Change
electrodes. Prepare patient’s skin.

ECG TOO LARGE

ECG signal too large for accurate shockable rhythm
analysis.

Reduce ECG size.

ECG V LEAD OFF

V LEAD cable is not properly attached.

Attach V LEAD cable.

ECG V1 LEAD OFF

Chest lead V1 is not properly attached to patient.

Reattach V lead. Check cable.

ECG V2 LEAD OFF

Chest lead V2 is not properly attached to patient.

Reattach V lead.

Check cable.

ECG V3 LEAD OFF

Chest lead V3 is not properly attached to patient.

Reattach V lead.

Check cable.

ECG V4 LEAD OFF

Chest lead V4 is not properly attached to patient.

Reattach V lead.

Check cable.

ECG V5 LEAD OFF

Chest lead V5 is not properly attached to patient.

Reattach V lead.

Check cable.

ECG V6 LEAD OFF

Chest lead V6 is not properly attached to patient.

Reattach V lead. Check cable.

ENTER ACCESS CODE

Manual mode access code needed.

Enter access code to enter manual mode with
AED.

ERASING REPORT

Summary report being erased.

FAX BUSY

FAX line at receiving FAX is busy.

FAX DIALING

Preparation for sending fax.

FAX DONE

Transmission complete.

FAX PREPARING

Preparing fax for transmission.

FAX SENDING

Transmitting fax.

Error Message

Explanation

User
Advisory

Technical Action

Advertising