Battery Well

15 Amp
Supply

?

Red

Black

Do this...

Verify that...

Pass/Fail

4.1

o o

4.2

4.3

o o

4.4

o o

4.5

o o

4.6

o o

E Series Service Manual

1-12

9650-1450-01 Rev. D

Turn the selector switch to MONITOR.

(For AED units, turn the selector switch to ON
and select Manual mode.)

The unit does not turn on.

Turn the selector switch to the OFF position.

Adjust the power supply voltage to 10.3V and
turn the selector switch to MONITOR (for AED
units, turn the selector switch to the ON
position).

The unit turns on.

Low Battery Test

Set voltage to 9.9V.

The unit does not display the LOW BATTERY message.

Set voltage to 9.4V.

The unit displays the LOW BATTERY message within 30 seconds.

Shut Down Voltage Test

Set voltage to 8.5V.

The unit shuts off within 30 seconds.

Record your results on the Maintenance Tests Checklist.