0 heart rate alarm test, Heart rate alarm test – ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 28

Advertising
background image

E Series Service Manual

1-18

9650-1450-01 Rev. D

10.0 Heart Rate Alarm Test

Tools Needed: QED 6 Defibrillator Analyzer

Do this...

Verify that...

Pass Fail/NA

10.1

Turn the selector switch to MONITOR mode.

(For AED units, turn the selector switch to ON
and select Manual mode.)

Connect the ECG leads to the QED 6
Defibrillator Analyzer

.

Set the QED 6 Analyzer

to 120 BPM and the defibrillator to lead II.

The monitor screen displays the Lead II message and the waveform for

NSR ECG at 120 BPM +/- 2.

o o o

10.2

Press ALARMS buttons.

The monitor displays the alarm menu.

o o o

10.3

Press SELECT PARAM softkey until ECG HR
displays.

The cursor scrolls through parameters.

o o o

10.4

Press INC> for state.

The cursor scrolls through ENABLE, AUTO and DISABLE.

o o o

10.5

Press DEC>for state.

The cursor scrolls through ENABLE, DISABLE, AND AUTO.

o o o

10.6

Press INC> until ENABLE displays.

ENABLE appears.

o o o

10.7

Then press the NEXT FIELD softkey to select
the heart rate limit.

The cursor scrolls to Low field.

o o o

10.8

Press Return, then press the ALARM
SUSPEND
button.

No alarm sounds.

o o o

10.9

Remove a lead wire from the QED 6 analyzer.

The alarm symbol flashes and the heart symbol stops flashing. The ECG LEAD
OFF alarm tone sounds. Recorder prints a stripchart showing a low heart rate,
if enabled.

o o o

10.10

Re-attach ECG Lead wire to the QED 6
Analyzer and hold the ALARM SUSPEND
button on unit for 4 seconds.

The unit displays the alarm symbol with an “X” through it.

The heart symbol flashes with each QRS wave.

o o o

10.11

Press the ALARM SUSPEND button to enable
the alarms.

The alarm symbol does not have an “X” through it.

o o o

10.12

Set QED 6 Analyzer to 160 BPM or higher.

The Heart Rate Value is highlighted, the alarm tone sounds, and the alarm and
the heart symbol both flash.

o o o

10.13

Press the ALARM SUSPEND button on the E
Series unit.

The alarm is suspended for 90 seconds. The unit displays the alarm symbol
with an “X” through it, and the heart symbol flashes with each QRS wave.

o o o