Connector panel removal, Connector panel removal -29 – ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 139

Advertising
background image

E Series Service Manual

9650-1450-01 Rev. D

5-29

Connector Panel Removal

Note:

There are 2 styles of SPO2 connector (MS-11 & MX-1) used on the E Series, use the appropriate instructions for disassembly.

1. (MS-11 Style) Remove the 2 screws securing the panel to the

housing.

2. Lift assembly out, gently sliding rubber ground sleeve off

mounting post.

3. Remove rubber gasket (black foam) from connector panel. ECG

and SPO2 connectors can now be removed.

Note:

During reassembly ensure conductive sleeve is in place next to SPO2
connector.