E Series Service Manual

9650-1450-01 Rev. D

5-11

NIBP Pump Removal

1. Cut tie wrap, and disconnect hose from the NIBP fitting.

2. Remove the screw that is holding the NIBP Pump.

Note:

When reinstalling, do not over tighten tie wrap.

3. Disconnect ECG flex cable from system board, by releasing

jams on each end of the connector, then gently pull straight out.

4. Unplug cable from system interconnect.