ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 46

Advertising
background image

E Series Service Manual

1-36

9650-1450-01 Rev. D

5

If you are using the Fluke® BiomedicalCuftlink Simulator,the volume leak reading should be <10.

6

If you are using the Fluke® BiomedicalCuftlink Simulator,the volume leak reading for a failure should be >10.

21.5

Press the START TEST softkey within 30
seconds after closing the valve.

Note: This test takes approximately 3 minutes.

After approximately 1 minute, a number appears in the upper middle area of
the NIBP simulator display.

If the simulator:

Displays a Volume Leak reading <5,

5

then the E Series unit has passed the

test.

Displays a Volume Leak reading >5,

6

then the E Series unit has failed the

test.

Displays no Volume Leak reading, but maintains a stable pressure reading
at or above 200 mmHg, then the E Series unit has passed the test; there is
no volume leak.

In addition, the E Series unit displays the simulator’s pressure reading in the
“Cuff Pressure” field.

After approximately 3 minutes, the valves open on the E Series unit.

o o

21.6

On the NIBP Analyzer, press the STOP TEST
softkey.

The NIBP simulator terminates the Volume Leak Test.

o o

21.7

On the E Series unit, press the EXIT softkey
twice.

The E Series unit returns to the main NIBP Service M

ode screen, then to

normal Monitor mode operation.

o o

Record your results on the Maintenance Tests Checklist.

Do this...

Verify that...

Pass/Fail