0 power supply test (optional), Power supply test (optional) – ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 21

Advertising
background image

E Series Service Manual

9650-1450-01 Rev. D

1-11

4.0 Power Supply Test (Optional)

Tools Needed:

1 red miniature alligator to miniature alligator test lead.

1 black miniature alligator to miniature alligator test lead.

DC power supply (15 Amp minimum).

0.1 ohm



resistor (

¼

W or greater).

1000 ohm 1% resistor (

¼

W or greater).

Fluke 75 multimeter or equivalent.

Test Setup:

1. Make sure the unit and power supply are turned off.

2.

Connect one end of the black lead to the “-” terminal in the battery well.

3.

Connect the other end of the black lead to the “-” terminal of the power supply.

4.

Connect the red lead to “+” terminal socket of the battery well. Use the middle pin with the plastic guard around it. Connect the other end of the red lead
to the “+” terminal of the power supply.

5.

Set the power supply voltage to 7V.

Caution

Be sure to connect the power supply properly to the E Series battery well terminals or damage to the unit may result. Do NOT raise the
power supply voltage above 12V.