9650-1450-01 Rev. D
1-11
4.0 Power Supply Test (Optional)
Tools Needed:
•
1 red miniature alligator to miniature alligator test lead.
•
1 black miniature alligator to miniature alligator test lead.
•
DC power supply (15 Amp minimum).
•
0.1 ohm
resistor (
¼
W or greater).
•
1000 ohm 1% resistor (
¼
W or greater).
•
Fluke 75 multimeter or equivalent.
Test Setup:
1. Make sure the unit and power supply are turned off.
2.
Connect one end of the black lead to the “-” terminal in the battery well.
3.
Connect the other end of the black lead to the “-” terminal of the power supply.
4.
Connect the red lead to “+” terminal socket of the battery well. Use the middle pin with the plastic guard around it. Connect the other end of the red lead
to the “+” terminal of the power supply.
5.
Set the power supply voltage to 7V.
Caution
Be sure to connect the power supply properly to the E Series battery well terminals or damage to the unit may result. Do NOT raise the
power supply voltage above 12V.