8.0 Calibrating Pulses on Strip Chart Test
Tools Needed: None
Test Setup: None
.
Do this...
Verify that...
Pass/Fail/NA
8.1
8.2
o o o
Press the RECORDER button.
Press and hold SIZE button to activate the
calibration signal.
The strip chart displays a signal of 300 ppm with an amplitude of
10 mm +/- 1 mm. The signal also appears on the video display.
Record your results on the Maintenance Tests Checklist.