8.0 Calibrating Pulses on Strip Chart Test

Tools Needed: None

Test Setup: None

.

Do this...

Verify that...

Pass/Fail/NA

8.1

8.2

o o o

Press the RECORDER button.

Press and hold SIZE button to activate the
calibration signal.

The strip chart displays a signal of 300 ppm with an amplitude of

10 mm +/- 1 mm. The signal also appears on the video display.

Record your results on the Maintenance Tests Checklist.