15.0 Advisory Message Test

Tools Needed: QED 6 Defibrillator Analyzer.

Test Setup:

1. Connect the universal cable via the adapter (D.N.I #3010-0378).

2.

Attach the E Series to the QED 6 Defibrillator Analyzer.

Do this...

Verify that...

Pass/Fail

15.1

15.2

o o

15.3

o o

15.4

o o

Turn the selector switch to DEFIB mode.

(For AED units, turn the selector switch to ON.)

Select VF (ventricular fibrillation) on the
simulator, then press the ANALYZE button.

The unit displays the following messages:

ANALYZING ECG.

STAND CLEAR.*

SHOCK ADVISED.*

PRESS SHOCK.*+

*AED’s audio prompts are standard. Advisory audio prompts are user
configurable.

+If configured for auto charge.

Press the SHOCK button.

The unit discharges.

Select the NSR (normal sinus rhythm) on the
simulator, then press the ANALYZE button.

The unit displays the following messages:

ANALYZING ECG.

STAND CLEAR.*

NO SHOCK ADVISED.*

*AED’s audio prompts are standard.

Record your results on the Maintenance Tests Checklist.