0 physical inspection of the unit, Physical inspection of the unit – ZOLL E Series Monitor Defibrillator Rev D User Manual
Page 15

E Series Service Manual
9650-1450-01 Rev. D
1-5
1.0 Physical Inspection of the Unit
Tools Needed: Battery.
Test Setup: None.
Observe this...
Pass/Fail/NA
1.1
Housing
Is the unit clean and undamaged?
o o
1.2
Does the unit show signs of excessive wear?
o o
1.3
Does the handle work properly?
o o
1.4
Does the recorder door open and close properly?
o o
1.5
Are input connectors clean and undamaged?
o o
1.6
Are there any cracks in the housing?
o o
1.7
Do the front panel or selector switches have any damage or cracks?
o o
1.8
Are there any loose housing parts?
o o
1.9
Cables
Are all cables free of cracks, cuts, exposed or broken wires?
o o
1.10
Are all cable bend/strain reliefs undamaged and free of excessive cable wear.
o o
1.11
Paddles
Do the adult and pedi plates have major scratches or show signs of damage?
o o o
1.12
Do the adult shoes slide on and off easily to expose the covered pedi plates?
o o o
1.13
Are the paddles clean (e.g., free of gel) and undamaged? (if applicable)
o o o
1.14
Battery
Place battery in battery well.
o o