0 barometric pressure calibration check, Barometric pressure calibration check – ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 42

Advertising
background image

E Series Service Manual

1-32

9650-1450-01 Rev. D

19.0 Barometric Pressure Calibration Check

Tools Needed: None.

Test Setup: None

Do this...

Verify that...

Pass/Fail/NA

19.1

19.2

o o o

19.3

o o o

19.4

o o o

19.5

o o o

*The barometric pressure can be obtained from a calibrated barometer, or from the National Weather Service at

www.nws.noaa.gov

(enter your local zip code to

get the local barometric pressure). Note that the barometric pressure is in inches of mercury, multiply it by 25.4 to convert to mmHg.

Obtain the local barometric pressure in mmHg.*

Turn the E Series unit off, and then wait

10 seconds. While pressing and holding the
second softkey from the left, turn the selector
switch to Monitor (ON for AED units).

The unit displays the EtCO

2

Calibration screen.

Wait for the sensor to warm up.

The unit displays the message WARM UP for approximately one minute.

Press the Restore Defaults softkey.

The difference between the current barometric pressure reading (obtained in
Step 19.1) and the barometric pressure reading that the E Series unit displays
is <30.

Use the Inc> and Dec< softkeys to enter the
current barometric pressure on the unit. When,
finished, press the RETURN key.

Record your results on the Maintenance Test Checklist.