E Series Service Manual

1-34

9650-1450-01 Rev. D

20.9

Attach a regulated flowing gas mixture of 5%
CO

2

/balance Nitrogen (N

2

) to the airway

adapter.

The gas flow rate should already be preset to 2 to 5 liters per minute.

20.10

Set the Gas Balance settings of the
CAPNOSTAT 5 CO

2

Sensor to that of the

calibration gas mixture (N

2

, N

2

O, or He).

The default gas balance is N

2

.

20.11

Allow a few seconds for the gas mixture to
stabilize and observe the CO

2

Percent value.

The expected value is 4.8 to 5.2.

o o o

20.12

Press the Return softkey to return to the main
EtCO

2

Calibration screen.

20.13

Turn the device off when calibration is
complete.

Record your results on the Maintenance Test Checklist.

Do this...

Observe this...

Pass/Fail/NA