E Series Service Manual
1-34
9650-1450-01 Rev. D
20.9
Attach a regulated flowing gas mixture of 5%
CO
2
/balance Nitrogen (N
2
) to the airway
adapter.
The gas flow rate should already be preset to 2 to 5 liters per minute.
20.10
Set the Gas Balance settings of the
CAPNOSTAT 5 CO
2
Sensor to that of the
calibration gas mixture (N
2
, N
2
O, or He).
The default gas balance is N
2
.
20.11
Allow a few seconds for the gas mixture to
stabilize and observe the CO
2
Percent value.
The expected value is 4.8 to 5.2.
o o o
20.12
Press the Return softkey to return to the main
EtCO
2
Calibration screen.
20.13
Turn the device off when calibration is
complete.
Record your results on the Maintenance Test Checklist.
Do this...
Observe this...
Pass/Fail/NA