ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 65

Advertising
background image

E Series Service Manual

9650-1450-01 Rev. D

2-13

SpO

2

COMM ERR

No transmissions from SpO

2

unit received. Communication

error or no communication from Sp0

2

module.

Replace Sp0

2

module and/or system board.

SpO

2

PULSE SEARCH

Pulse search in progress.

STAND CLEAR

(Auto defib mode only.) Single analysis mode just turned
on and defib idle. Patient rhythm is being analyzed.

SYSTEM FAULT 1

No Watchdog.

Turn off unit and then turn on to reset. If fault
persists, replace system board.

SYSTEM FAULT 6

No communications received from ECU for 4 seconds.

Turn off unit and then turn on to reset. If fault
persists, replace system board.

SYSTEM FAULT 7

The A/D converter is not performing conversions in a
timely manner.

Replace system board.

SYSTEM FAULT 36

PS_MON is out of range. Pace/defib is disabled as long as
condition exits.

Replace system board.

SYSTEM FAULT 37

Disable pace/defib and MFE monitoring.

Replace system board.

SYSTEM FAULT 38

Failure to shutdown after “shutdown order” is written to the
RTC.

Replace system board.

TEST FAILED

MCU performed ipeak test (defib peak current) and unit
failed during 30J self test.

Replace universal cable, paddles or high voltage
module, capacitor, or system board.

TEST OK

MCU performed ipeak test (defib peak current) and unit
passed 30J self test.

USE PADDLE DISCHG

Front Panel discharge button is pressed when either
external paddles or internal spoons with discharge buttons
are connected.

USE PADS

(AUTO DEFIB MODE ONLY.)

Attempt to defib with paddles in auto defib (AED) mode.
Defib only allowed using PADS in AED modes.

USE PADS TO PACE

External paddles detected in pace mode.

Error Message

Explanation

User
Advisory

Technical Action