O o o o – ZOLL E Series Monitor Defibrillator Rev D User Manual
Page 40

E Series Service Manual
1-30
9650-1450-01 Rev. D
17.5
Using the Index SpO
2
Simulator, press the
BPM- softkey until the heart rate is 50 BPM
The SpO
2
simulator screen displays an SpO
2
saturation of 96-100%.
The E Series monitor displays a heart rate of 46-54 BPM.
o o
o o
17.6
Using the Index SpO
2
Simulator, press the 02+
softkey until the SpO
2
output is at 72%.
The SpO
2
simulator screen displays an SpO
2
saturation of 70-74%.
The E Series monitor displays a heart rate of 46-54 BPM.
o o
o o
17.7
Press Wave 2 softkey.
Select the SpO2 waveform.
Plethysmographic waveform appears on the ZOLL display.
o o
17.8
Press RECORDER.
The plethysmographic waveform prints on the strip chart paper.
o o
17.9
Using the Index SpO
2
Simulator, press the
BPM- softkey until the heart rate is at 230 BPM.
The SpO
2
simulator screen displays an SpO
2
saturation rate of 70-74%.
The E Series monitor displays a heart rate in the heart position of
226-234 BPM.
o o
17.10
Remove the Masimo
patient cable.
Record your results on the Maintenance Tests Checklist.
Do this...
Verify that...
Pass/Fail