O o o o – ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 40

Advertising
background image

E Series Service Manual

1-30

9650-1450-01 Rev. D

17.5

Using the Index SpO

2

Simulator, press the

BPM- softkey until the heart rate is 50 BPM

The SpO

2

simulator screen displays an SpO

2

saturation of 96-100%.

The E Series monitor displays a heart rate of 46-54 BPM.

o o
o o

17.6

Using the Index SpO

2

Simulator, press the 02+

softkey until the SpO

2

output is at 72%.

The SpO

2

simulator screen displays an SpO

2

saturation of 70-74%.

The E Series monitor displays a heart rate of 46-54 BPM.

o o
o o

17.7

Press Wave 2 softkey.

Select the SpO2 waveform.

Plethysmographic waveform appears on the ZOLL display.

o o

17.8

Press RECORDER.

The plethysmographic waveform prints on the strip chart paper.

o o

17.9

Using the Index SpO

2

Simulator, press the

BPM- softkey until the heart rate is at 230 BPM.

The SpO

2

simulator screen displays an SpO

2

saturation rate of 70-74%.

The E Series monitor displays a heart rate in the heart position of

226-234 BPM.

o o

17.10

Remove the Masimo

patient cable.

Record your results on the Maintenance Tests Checklist.

Do this...

Verify that...

Pass/Fail