System board removal, System board removal -13 – ZOLL E Series Monitor Defibrillator Rev D User Manual
Page 123
Advertising
E Series Service Manual
9650-1450-01 Rev. D
5-13
System Board Removal
1. Remove plastic spacer by gently lifting.
2. Remove PCMCIA short connecting ribbon cable from both
sides.
Note:
During reassembly verify the spacer ridge sits on board.
3. Remove flat UBP cable from both ends by releasing jams on
each end of the connector, then pull straight out.
4. Gently remove the 2 flat cables by releasing jams on each
end of the connector, then gently pull straight out.