14.0 Summary Report Test

Tools Needed: QED 6 Defibrillator Analyzer.

Test Setup:

1. Connect the universal cable to the QED 6 analyzer.

2.

If you are using paddles, place the paddles on the analyzer’s discharge plates.

Do this...

Verify that...

Pass/Fail

14.1

o o

14.2

o o

14.3

o o

14.4

o o

Press and hold the SUMMARY softkey for 4 to
8 seconds to display erase options. Press the
ERASE SUMMARY softkey to erase any
previously stored data erase any previously
stored data.

The unit displays the message, ERASING REPORT.

Set selector switch to DEFIB. Select 200J using
the ENERGY SELECT button, and press the
CHARGE
button. When charged, press the
SHOCK
button to discharge into the defibrillator
analyzer.

The unit successfully discharges and prints a strip chart.

Wait 18 seconds, then press the Code Marker
softkey. Press the CPR softkey.

The Code Markers appear.

Turn the unit off. Wait 10 seconds and then turn
the unit on. Press the SUMMARY softkey, then
press the PRINT CHART softkey.

The summary report prints. The report displays the correct date, time, the
shock delivered and Code Marker event.

Record your results on the Maintenance Tests Checklist.