ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 38

Advertising
background image

E Series Service Manual

1-28

9650-1450-01 Rev. D

16.6

Decrease the output to 40mA.

Output on the QED 6 analyzer is 40mA or +/- 5mA. Pulse width is

40mS +/-2mS.

o o

16.7

Increase the output to 60mA.

Decrease the rate to 30 ppm.

Pacer rate on QED 6 analyzer is 29-31 ppm

o o

16.8

Increase the rate to 180ppm.

Pacer rate on QED 6 analyzer is 177-183 ppm

o o

16.9

Connect the ECG cable to the E Series and
QED 6 analyzer. Select the ECG at 60 BPM on
the QED 6 analyzer. Decrease the pacer rate
on the unit to 58ppm.

The unit displays ECG at 60 BPM with no stimulus markers.

o o

16.10

Press the Async Pace softkey.

The unit displays ECG at 60 BPM with stimulus markers, and displays the
Async pace message.

o o

16.11

Turn off the QED 6 analyzer. Set Pacer Rate to
100ppm. Press the RECORDER ON button.

The unit displays pace stimulus markers every 15mm +/-1mm.

o o

16.12

Press and hold 4:1 button.

The unit displays pace stimulus markers every 60 mm+/- 1.5 mm.

o o

Record your results on the Maintenance Tests Checklist.

Do this...

Verify that...

Pass/Fail