ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 50

Advertising
background image

Do this...

Verify that...

Pass/Fail

23.1

23.2

o o

23.3

Systolic pressure (115 - 125 mmHg)

Diastolic pressure (75 - 85 mmHg)

Mean pressure (88 - 98 mmHg)

o o

23.4

o o

E Series Service Manual

1-40

9650-1450-01 Rev. D

4

These values only apply for test set-ups utilizing the BP Pump 2 Simulator. Variations of the test set-up or different simulators may produce readings outside

the provided values and will require end-user facility to establish the appropriate offset and tolerances. Please contact ZOLL Technical Support i f you require
assistance establishing the offset o f y o u r simulator and test set-up.

Turn the selector switch to MONITOR mode.

(For AED units, turn the selector switch to ON
and select Manual mode.)

Ensure that the LEADS parameter is set to
PADS (default).

If necessary, press the LEADS button to cycle
through the values to select PADS.

The E Series unit displays PADS in the Lead selection field on the monitor.

Press the NIBP button on the E Series front
panel.

The E Series displays the following measurements:

Press the PARAM softkey, then press the
TREND softkey, followed by the NIBP Trend
softkey.

The E Series unit displays a summary of the NIBP measurements, including
the pulse rate reading (in the range of 77 - 83 bpm).

Record your results on the Maintenance Tests Checklist.

Advertising