Cpr monitoring – ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 250

Advertising
background image

E Series Operator’s Guide

A-4

9650-1210-01 Rev. R

CPR Monitoring

Heart Rate Alarm

Screen icon indicates activated/deactivated status. User-selectable.
Tachycardia 60 bpm - 280 bpm, bradycardia 20 bpm - 100 bpm

Tall T-wave
Rejection

1.0 mV

Heart Rate
Averaging

The E Series averages the interval between the last 5 detected beats. On startup, the
E Series averages the rate between detected beats once two beats are detected, until a full 5
beats have been received. The rate is updated every beat. After this condition is met, the
meter is updated every beat with an average of the last 5 beats.

If a period of time greater than 5 seconds elapses without a beat detected, the heart rate
meter reports a rate of 0 bpm, which is repeated every 5 seconds.

Accuracy and
Response Time to
Irregular Rhythm

Averaging over 5 R-R intervals, per AAMI EC 13:2002:

Ventricular Bigeminy (Figure 3a) - 40 bpm

Slow alternating ventricular bigeminy (Figure 3b) - 60 bpm

Rapid alternating ventricular bigeminy (Figure 3c) - 120 bpm

Bidirectional systoles (Figure 3d) - 90 bpm

Response Time to
Change in Heart
Rate

80 to 120 bpm: 4 seconds
80 to 40 bpm: 4 seconds

Time to Alarm for
Tachycardia

206 bpm (1 mV): 4.80 seconds
206 bpm (halved amplitude): 9.90 seconds
206 bpm (doubled amplitude): 4.00 seconds
195 bpm (2 mV): 4.20 seconds
195 bpm (halved amplitude): 8.00 seconds
195 bpm (doubled amplitude): 4.00 seconds

Leads Off Sensing

A DC current of 0.04

A per lead wire is supplied to the patient.

Active Noise
Suppression

The sum of all leadwire currents is returned via the active noise suppression leadwire:

0.08

A DC in 3 lead mode

0.16

A DC in 5 lead mode

0.36

A DC in 12 lead mode

1 Volt ECG Out

1.0 Volt/cm (volt per centimeter) of deflection on stripchart recorder
< 25 ms delay from patient ECG input

Bandwidth

150 Hz

Output impedance = 250 ohms
Internal pacemaker pulses represented as on display

ECG x 1000 provided at rear connector panel via a standard 3.5 mm mono jack. ECG signal
is on tip; ground is on ring

Display Format

Non-fade moving bar display

Compression Depth

0.75 to 3 inches ±0.25 inches
1.9 to 7.6 cm ±0.6 cm

Compression Rate

50 to 150 compressions per minute