5.0 Leakage Current Test
Tools Needed: See the manufacturer’s instructions or supplied specifications for the leakage tester you use.
Test Setup: See the manufacturer’s instructions or supplied specifications for the leakage tester you use. Repeat leakage test with accessories: MFC,
external paddles, and anterior/posterior paddles.
Maximum Leakage Acceptance Limits
*Single fault considered AC mains on applied part.
Normal Condition
Single Fault Condition*
ECG
10
50
MFC
100
100
Earth
500
1000