E Series Service Manual

9650-1450-01 Rev. D

1-13

5.0 Leakage Current Test

Tools Needed: See the manufacturer’s instructions or supplied specifications for the leakage tester you use.

Test Setup: See the manufacturer’s instructions or supplied specifications for the leakage tester you use. Repeat leakage test with accessories: MFC,
external paddles, and anterior/posterior paddles.

Maximum Leakage Acceptance Limits

*Single fault considered AC mains on applied part.

Normal Condition

Single Fault Condition*

ECG

10



50



MFC

100



100



Earth

500



1000

