Resp system message – ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 106

Advertising
background image

Chapter 7 Monitoring Respiration (Resp) and Heart Rate (HR)

7-10

www.zoll.com

9650-001355-01 Rev. H

RESP System Message

When monitoring Respiration using impedance pneumography, the X Series unit may display
the following message:

System Message

Cause

RESPIRATION FAULT

The impedance measured through the ECG lead is
too high for the unit to detect Respiration changes,
but not high enough to indicate an ECG lead fault
(the impedance may vary between 1800 and 2800
ohms).

Advertising