ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 85

Advertising
background image

Alarm Options

9650-001355-01 Rev. H

X Series Operator’s Guide

5-9

SpO

2

(%)

Entire range

Limit = 100 (Adult and
Pediatric)
Limit = Numeric + 5 (Neo-
nate)

Limit = Numeric – 5

SpCO

(%)

Entire range

Limit = Numeric + 2
Upper limit 40

Limit = Numeric - 2
Lower limit 0

SpMet

(%)

Entire range

Limit = Numeric + 2
Upper limit 15

Limit = Numeric - 2
Lower limit 0

EtCO

2

(mmHg)

Entire range

Limit = Numeric + 10

Limit = Numeric - 5 mmHg
unless the numeric falls
below the lower alarm limit
range, in which case Stat
Set sets the lower limit to
15 mmHg.

FiCO

2

(mmHg)

Entire range

Limit = Numeric + 5

N/A

Temp

(°C)

Entire range

Limit = Numeric + 0.5

Limit = Numeric – 0.5

Temp

(°F)

Entire range

Limit = Numeric + 0.9

Limit = Numeric – 0.9