Alarm Options
9650-001355-01 Rev. H
X Series Operator’s Guide
5-9
SpO
2
(%)
Entire range
Limit = 100 (Adult and
Pediatric)
Limit = Numeric + 5 (Neo-
nate)
Limit = Numeric – 5
SpCO
(%)
Entire range
Limit = Numeric + 2
Upper limit 40
Limit = Numeric - 2
Lower limit 0
SpMet
(%)
Entire range
Limit = Numeric + 2
Upper limit 15
Limit = Numeric - 2
Lower limit 0
EtCO
2
(mmHg)
Entire range
Limit = Numeric + 10
Limit = Numeric - 5 mmHg
unless the numeric falls
below the lower alarm limit
range, in which case Stat
Set sets the lower limit to
15 mmHg.
FiCO
2
(mmHg)
Entire range
Limit = Numeric + 5
N/A
Temp
(°C)
Entire range
Limit = Numeric + 0.5
Limit = Numeric – 0.5
Temp
(°F)
Entire range
Limit = Numeric + 0.9
Limit = Numeric – 0.9