Configuring Respiration (RR/BR) Alarms and Settings

9650-001355-01 Rev. H

X Series Operator’s Guide

7-3

Configuring Respiration (RR/BR) Alarms and Settings

The X Series unit allows you to enable and disable the Respiration (

RR/BR

) Rate alarm, to set

alarm limits, and to specify the ECG monitoring source for the Respiration rate.

Enabling/Disabling RR/BR Alarms and Setting Alarm Limits

When enabled, the X Series unit sounds alarms whenever the patient’s respiration rate is above
or below the specified respiration rate alarm limits.

To enable (or disable) Resp alarms and set Upper and Lower alarm limits, you can either do so

through the Alarms quick access key (

), or the Resp Parameter Control Panel.

To configure

RR/BR

alarms through the Alarms quick access key:

1. Press .

2.

Press .

3.

Press the Limits quick access key (

). Use the navigation keys to highlight and select the

RR/BR Alarm

menu selection.

4.

On the

RR/BR

Alarm Settings menu, use the navigation keys to select the fields that you want

to change:

Upper Limit Enable

Lower Limit Enable

Upper Limit

Lower Limit

5.

When you are finished changing values on the alarm menu, navigate to the Backarrow key
to confirm your choices and exit the menu.