ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 278

Advertising
background image

C

HAPTER

24 M

AINTENANCE

24-4

www.zoll.com

9650-001355-01 Rev. H

8

Press the Select Energy arrow (up or down).

Verify that the defibrillation window displays
and the following message appears:

PADS SHORT DETECTED

SELECT 30J TO TEST

9

Press the Select Energy arrow (up or down)
to select 30 Joules on the X Series unit.

Verify that the defibrillator window shows
30 J as the selected energy.

10 Press the Charge button on the front panel.

Verify that a periodic tone sounds while the
defibrillator is charging. At the completion of
the charge cycle, the tone is continuous. The
charge time should be less than 7 seconds.

Verify that the Shock button (

) on the

front panel illuminates when the defibrillator
is charged.

11

Press and hold

on the front panel until

the unit discharges.

Verify that the defibrillator discharges and
that a DEFIB SHORT TEST PASSED
message displays.

If the message DEFIB SHORT TEST FAILED
displays, contact the appropriate technical
personnel or the ZOLL Technical Service
Department.

12 After the defibrillator display panel closes,

press the Pacer button on the front panel to
display the Pacer Settings dialog box.

Verify that the Pacer Settings dialog displays
with the pacer status bar behind it.

13 Perform the following steps:

Set the Pacer mode to Fixed.

Set the Pacer Rate to 60 ppm.

Set the Output to 100 mA.

Select the Start Pacer button.

Verify that the green pace indicator on the
pacer status bar flashes once each second.

Verify that pace markers are printed every 25
mm on the chart recorder printout.

Note: If a printout does not automatically
begin after starting the pacer, select the Print
quick access key to start a continuous
printout. Press the Print quick access key
again to stop the printout.

14 Press the Pacer button on the front panel

button again to close the Pacer Settings
dialog box.

Verify that the pacer status bar indicates
PACING: SHORT DETECTED.

Verify that the green pace indicator continues
to flash once each second.

Verify that no pacer-related error messages
display.

15 Disconnect the test connector from the MFC.

Verify that the pacer status bar displays:

PACING: CHECK THERAPY ELECTRODES

16 Press the Pacer button on the front panel to

display the Pacer Settings dialog. Select Turn
Pacer Off
to turn off the pacer.

Verify that the Pacer Settings dialog is closed
and that the pacer status bar is removed from
the display.

17 Reconnect the external AC Adapter to the X

Series rear panel.

Verify that the green Auxiliary Power indicator
light on the X Series front panel is
illuminated.

Function

Response

Advertising