Pacing in Demand Mode

9650-001355-01 Rev. H

X Series Operator’s Guide

18–3

2 Apply ECG Electrodes/Hands-Free Therapy Electrodes

Apply ECG electrodes, attach lead wires, and connect the ECG cable to the X Series side panel
(see Chapter 6, "Monitoring ECG" for instructions on attaching ECG electrodes to the patient).
Attach hands-free therapy electrodes according to instructions on the electrode packaging.
Connect these therapy electrodes to the multifunction cable (MFC).

Note:

Do not use ZOLL OneStep cables for pacing.

Therapy Electrode Application

WARNING!

Poor adherence and/or air under the therapy electrodes can lead to the possibility of
arcing and skin burns.

1. Apply one edge of the pad securely to the patient.

2.

Roll the pad smoothly from the applied edge to the other, being careful not to trap any air
pockets between the gel and skin.

3.

Ensure that hands-free therapy electrodes are making good contact with the patient’s skin and
are not covering any part of any other ECG electrodes.

4.

Press the Lead quick access key and select I, II, or III to provide the largest amplitude QRS
complex.

Note:

When the Pacer is on, the lead selection is restricted to Leads I, II, or III.

5.

Verify that R-waves are being properly detected by confirming that a QRS tone occurs with
each displayed R wave or by verifying that the X Series unit’s heart rate display accurately
reflects the patient’s pulse rate.

3 Press Pacer button

Press the Pacer button on the front panel of the unit. The Pacer Settings window displays.

Pad

Skin

Pacer Settings

Mode

Rate

Output

Start Pacer

Turn Pacer Off

Demand

60

70