IBP System Messages
9650-001355-01 Rev. H
X Series Operator’s Guide
11-9
IBP System Messages
The X Series unit may display the following messages when monitoring IBP:
System Message
Cause
TRANSDUCER FAILURE
The IBP probe is damaged and needs to be replaced.
INCOMPATIBLE TRANSDUCER
The IBP probe is not compatible. See the Appendix B,
Accessories, for a list of ZOLL-approved IBP probes.
CHECK PROBE
The IBP probed has become disconnected.
ZERO PROBE
The IBP probe is connected and needs to be zeroed.
ZEROING
The IBP probe is zeroing.
IBP DISABLED
A system error has occurred, and the X Series unit should
be serviced.
ZERO REJECTED
The IBP probe was not removed due to a pulsatile
pressure signal, excessive IBP artifact or excessive
transducer offset.