Ibp system messages – ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 157

Advertising
background image

IBP System Messages

9650-001355-01 Rev. H

X Series Operator’s Guide

11-9

IBP System Messages

The X Series unit may display the following messages when monitoring IBP:

System Message

Cause

TRANSDUCER FAILURE

The IBP probe is damaged and needs to be replaced.

INCOMPATIBLE TRANSDUCER

The IBP probe is not compatible. See the Appendix B,
Accessories, for a list of ZOLL-approved IBP probes.

CHECK PROBE

The IBP probed has become disconnected.

ZERO PROBE

The IBP probe is connected and needs to be zeroed.

ZEROING

The IBP probe is zeroing.

IBP DISABLED

A system error has occurred, and the X Series unit should
be serviced.

ZERO REJECTED

The IBP probe was not removed due to a pulsatile
pressure signal, excessive IBP artifact or excessive
transducer offset.