Electrodes

The X Series units will defibrillate, cardiovert, and monitor ECG using hands-free therapy
electrodes. The X Series unit will pace using ZOLL hands-free therapy electrodes.

Energy Select, Charge and Shock controls are located on the paddles and front panel. When
using hands-free therapy electrodes, you must use the controls on the front panel of the unit. To
switch between paddles and hands-free therapy electrodes, remove the multifunction cable
(MFC or OneStep) from the apex paddle and connect the hands-free therapy electrodes to the
cable.

You should always check the expiration date on the electrode packaging. Do not use expired
electrodes, which might result in false patient impedance readings and affect the level of
delivered energy, or cause burns.

Note:

ZOLL electrodes contain no hazardous materials and may be disposed of in general
trash unless contaminated with pathogens. Use appropriate precautions when
disposing of contaminated electrodes.

When the patient is less than 8 years old or weighs less than 55 lb. (25 kg), use ZOLL OneStep
Pediatric defibrillation electrodes. Do not delay therapy to determine the patient’s exact age or
weight.

Batteries

X Series models use an easily replaced rechargeable lithium-ion battery pack (the SurePower II
Battery Pack). A new, fully charged battery pack typically delivers more than 6 hours of ECG
monitoring. Use of other functions (such as the defibrillator, printer, or pacemaker) reduces this
time.

When a LOW BATTERY icon appears on the display and the unit emits three beeps in
conjunction with the displayed battery icon, the battery must be replaced and recharged.

You can charge the battery by either of the following methods:

Internal charging — plug the X Series into an auxiliary power adapter to automatically
begin charging the installed battery pack. The front panel battery indicator operates as
follows:

Note:

Upon power up, it takes approximately 45 seconds for the LEDs on the battery to
accurately display run time.

This symbol on the electrode package is accompanied by the expiration date.

For Stat-padz

®

II, this symbol does not appear; the expiration date appears on the

lower right corner of the label, below the lot number.

When the indicator is:

It means:

Steady yellow

Battery is charging.

Steady green

Battery is charged.

Alternating yellow and
green

The charge state cannot be
determined or a battery charging
fault has been detected.

Not lit

No battery in device.