Alarms – ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 302

Advertising
background image

Specifications

A-16

www.zoll.com

9650-001355-01 Rev. H

Alarms

Heart Rate Alarms:

Audible: 5 pulse, 900 Hz tone, with a PW of 125 msec, a PRI of 250 msec, and a repetition
interval of 15 seconds.

Visual: Heart Rate Alarm causes the heart rate to be displayed in red, with a white background.

The red device status LED will flash a rate of 1.7 Hz.

Lead Fault Alarm:

Audible: 3 pulse, 500 Hz, triplet tone with a PW of 200 msec, a PRI of 200 msec. The lead
fault tone repeats at a repetition interval of 30 seconds.

Visual: Lead Fault condition causes a LEAD FAULT message to be displayed on the trace
along with a dashed line the width of the trace.

Physiological Alarms (NIBP, SpO

2

, Resp, CO

2

, IP & Temp):

Audible: Same as Heart Rate Alarm

Visual: Physiological alarms cause the alarming parameter to be displayed in Red with a white
background. The red device status LED will flash at a rate of 1.7 Hz.

Silence Duration: 90 seconds.

Invalid Operation Alert Tone:

A short, low-pitched tone is audible when a selected control button is unavailable for use or an
invalid entry is detected. Tone frequency is 160 Hz. Duration is 250 msec.

Delays (hold off time):

Heart rate/pulse rate:

i

f source is ECG, 5 seconds

if source is SpO

2

, 10 seconds

if source is IBP, 5 seconds

if source is NIBP, no hold off

SpO

2

, SpCO, and SpMet Saturation: 10 seconds

EtCO

2

: 7 seconds

FiCO

2

: 5 seconds

IBP (Systolic, Diastolic, Mean): 2 seconds

Temperature: 2 seconds

Advertising