Trends
The X Series unit accumulates a patient’s trend information by logging all monitored vital sign
measurements to memory at a user-configurable interval. It also logs all monitored vital sign
measurements when the following occurs:
•
An NIBP measurement is taken and the Trend on NIBP option is on
•
You press the Snapshot button (
) on the front panel
•
A patient alarm occurs and the Trend on Alarm option is on
The X Series unit can store at least 24 hours of trend information when logged at a 1 minute
trend interval. You can view, print, or save to external memory all logged trend information.
Displaying the Trends Status Window
The X Series unit displays the logged trend information in the Trends status window. Press the
Display/Home button (
) to display the Trends window, the primary ECG trace, and the
small numeric displays for each monitoring function: