Trends

The X Series unit accumulates a patient’s trend information by logging all monitored vital sign
measurements to memory at a user-configurable interval. It also logs all monitored vital sign
measurements when the following occurs:

An NIBP measurement is taken and the Trend on NIBP option is on

You press the Snapshot button (

) on the front panel

A patient alarm occurs and the Trend on Alarm option is on

The X Series unit can store at least 24 hours of trend information when logged at a 1 minute
trend interval. You can view, print, or save to external memory all logged trend information.

Displaying the Trends Status Window

The X Series unit displays the logged trend information in the Trends status window. Press the

Display/Home button (

) to display the Trends window, the primary ECG trace, and the

small numeric displays for each monitoring function: