Configuring the waveform display – ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 69

Advertising
background image

Configuring the Waveform Display

9650-001355-01 Rev. H

X Series Operator’s Guide

3-7

Configuring the Waveform Display

You can display up to four waveform traces on the Waveform Display window. The first
waveform trace always uses an ECG lead as its source (such as

Pads

or Leads

l

,

ll

, or

lll

, and so

on. The default is

Pads

). If

Pads

are not connected, the unit can be configured to

automatically default to another ECG lead for the first trace. As you insert the remaining three
traces, you can specify that the traces use an ECG lead as the waveform source, or that the trace
derive its waveform from other available monitoring functions (such as

Resp

,

CO2

,

SpO2

or

IBP channels

P1

,

P2

, or

P3

).

If configured, the unit can display four ECG traces on startup, when no other monitoring
devices are attached.

The X Series unit can also cascade a trace onto the adjoining trace area to double the duration
of the trace display.

On the Waveform Display window, to insert a new trace (

Insert

) or cascade (

Cascade

) a

displayed trace, highlight and select the trace label above the trace. In the following example,
the unit is configured to cascade the ECG Lead l trace:

Note:

The X Series unit automatically inserts a new waveform when you turn on a parameter
(CO

2

) or a new sensor signal is present (SPO

2

, IBP). The X Series unit automatically

removes a waveform when you turn off a parameter or remove a sensor and the unit
displays the resulting equipment alert.

06/06/2011 12:34:56

Adult

00:17:43

I, II,
III...

12

CO

2

R

SYNC

1 cm/mV

II

HR bpm

80

121

79

(96)

NIBP mmHg

38

12

BR

CO2 mmHg SpO2 %

97

T1 ºF

98.6

Pads

III

aVL
aVF

V

Insert

Source

II

aVR

I

Cascade

Advertising