Setting up Communications in the Supervisor Menu

9650-001355-01 Rev. H

X Series Operator’s Guide

22–17

Select Load from USB. A list of files on the USB device will appear.

Filenames can only be 24 characters long. If the files on the USB device are more than 24
characters, then they will not appear on the X Series unit screen.

Note:

The X Series unit only supports PFX format client certificates.

Delete a file by selecting Delete. After deleting a file, be sure to review the defined Wi-Fi
profiles and update any which referred to the deleted file.

To load a file, use the arrow keys to select the desired file. Only one file can be selected at a
time.

Select Load Selected. If a new USB device has been inserted, select Refresh List.

Note:

No message will appear to indicate that the file has loaded. After pressing Load

Selected, press the back arrow (

) to view the list of all currently loaded certificates

and keys.