Life Threatening Rhythm Alarms
When LTA monitoring is enabled, the X Series will monitor for the following life threatening
ECG rhythms: asystole, ventricular fibrillation, ventricular tachycardia, extreme bradycardia,
and extreme tachycardia.
Depending upon configuration of the unit, it generates visible and audible alarms as needed.
Note:
LTA monitoring is not available in AED mode.
Equipment Alert Display
When a problem with the X Series unit or an attached sensor triggers an alert, in addition to
sounding an equipment alert, the X Series unit displays an alert message (yellow background,
black text).
Warning!
Always respond immediately to a system alarm since the patient may not be monitored
during certain alert conditions.
In the following example, an equipment alert message indicates that the SpO
2
sensor has
become unattached (
SpO2 Check Sensor
) from the unit: