Setting CO

2

and Respiration Rate Alarms

The X Series unit sounds alarms whenever measurements are outside set limits for the
following:

High and Low EtCO

2

High and Low Respiration Rate (in Breaths/Minute)

High FiCO

2

No Breath Alarm

Note:

Use the Resp Parameter Control Panel to set the duration of the No Breath Alarm, or to
disable it (see “Using the Resp Parameter Control Panel” on page 7-4).

Enabling/Disabling Alarms and Setting CO

2

Alarm Limits

You can enable (or disable) CO

2

alarms and set upper and lower alarm limits through the

Alarms quick access key (

) or the CO2 Parameter Control Panel.

To configure CO

2

alarms through the Alarms quick access key:

1. Press the More quick access key (

).

2.

Press .

3.

Press the Limits quick access key (

).

4.

Use the navigation buttons to select appropriate alarm menu. For CO

2

, the alarm menus are:

EtCO2 Alarm, FiCO2 Alarm, or RR/BR Alarm.

5.

On the alarm settings menu, use the navigation keys to select the fields that you want to
change. The fields are

Upper Limit Enable

Lower Limit Enable (only EtCO2 and RR/BR Alarms)

Upper Limit

Lower Limit (only EtCO2 and RR/BR Alarms)

6.

When you are finished changing values on the alarm menu, select the Backarrow key to exit
the menu.

Limits