ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 173

Advertising
background image

AED Operation

9650-001355-01 Rev. H

X Series Operator’s Guide

13–9

CHECK PULSE

If configured to do so, this message is displayed and announced in the following situations:

After a No Shock Advised analysis result

During the CPR interval after a No Shock Advised analysis result

After the final shock is delivered

IF NO PULSE, PERFORM CPR

If configured to do so, this message is displayed and announced in the following situations:

During the CPR interval after a No Shock Advised analysis result

During the start of extra CPR interval

PERFORM CPR

If configured to do so, this message is displayed and announced during the CPR interval after a
No Shock Advised analysis result.

STOP CPR

After performing CPR for the configured period, the unit will issue a STOP CPR prompt while
it restarts ECG analysis.

PUSH HARDER

This message is announced when the chest compressions applied during CPR are not deep
enough.

GOOD COMPRESSIONS

This message is announced when the chest compressions applied during CPR are deep enough.

CHECK PADS

This message is displayed and announced when the therapy pads have been disconnected from
the patient.

CHECK PATIENT

This message is displayed and announced when the unit is paused and detects a shockable
rhythm during continuous background ECG analysis. The prompt persists as long as a
shockable rhythm is being detected. Press the ANALYZE button to resume ECG analysis.