Pacing, Fixed – ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 223

Advertising
background image

Pacing in Fixed Mode

9650-001355-01 Rev. H

X Series Operator’s Guide

18–7

3 Press Pacer button

Press the Pacer button on the front panel of the unit. The Pacer Settings window displays.

4 Set Mode

Use the arrow keys to navigate to Mode, press the Select button, then use the arrow keys and
Select button to set the Pacer Mode to Fixed. The selected mode is displayed in the pacing
window.

Note:

The initial pacing mode can be set in Defib/Pacer Default settings in the Supervisor
Setup menu.

5 Set Pacer Rate

Use the arrow keys to navigate to Rate, press the Select button, then use the arrow keys and the
Select button to set the Pacer Rate to a value 10-20 ppm higher than the patient’s intrinsic heart
rate. If no intrinsic rate exists, use 100 ppm. You can increase or decrease the pacer rate by a
value of 5 ppm for rates below 100, and by 10 ppm for rates above 100.

Note:

The initial pacing rate can be set in Defib/Pacer Default settings in the Supervisor
Setup menu.

6 Turn On Pacer

Use the arrow keys to navigate to Start Pacer, and press the Select button to select it. The
Pacing window displays behind the Pacer Settings window.

7 Set Pacer Output

In the Pacer Settings window, use the arrow keys and the Select button to adjust the pacer
output. The pacer output is adjustable in 10 mA increments when increasing the output, and in

Pacer Settings

Mode

Rate

Output

Start Pacer

Turn Pacer Off

Demand

60

70

PACING

60

ppm

Rate

70

mA

Output

Fixed

Mode