Applying electrodes to the patient – ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 179

12-Lead ECG Monitoring Setup
9650-001355-01 Rev. H
X Series Operator’s Guide
14-5
Applying Electrodes to the Patient
Depending on local usage, ECG lead wires are marked with certain labels. Refer to the
following table for labels and color codes for the different lead sets
1
American Heart Association
2
International Electrotechnical Commission
Patients should be in a resting, supine position when performing 12-Lead ECG monitoring.
ZOLL Medical Corporation recommends placing the limb electrodes anywhere along the
ankles and wrists.
When it is difficult for the patient to remain motionless due to shivering, muscle tremors, or
transport vehicle movement, place limb electrodes on the patient’s thorax for better results.
(Refer to the following two diagrams for limb electrode placement).
Location
AHA
1
Labels
IEC
2
Labels
Right Arm
RA (white)
R (red)
Left Arm
LA (black)
L (yellow)
Right Leg
RL (green)
N (black)
Left Leg
LL (red)
F (green)
Chest
V1
C1
Chest
V2
C2
Chest
V3
C3
Chest
V4
C4
Chest
V5
C5
Chest
V6
C6