Pediatric pacing – ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 225

Advertising
background image

Pacing in Fixed Mode

9650-001355-01 Rev. H

X Series Operator’s Guide

18–9

Pediatric Pacing

Noninvasive pacing of pediatric patients is performed in an identical manner to adult pacing.
Smaller size pediatric therapy electrodes are available for patients weighing less than 33 lbs/15
kg. If it is necessary to pace for more than 30 minutes, periodic inspection of the underlying
skin is strongly advised. Carefully follow all instructions on electrode packaging.

Pace Fault

The X Series unit may display the following messages when pacing.

System Message

Description

PAUSED

The pacer has paused pacing the patient.

PACING

The pacer is pacing the patient.

PACING: CHECK THERAPY
ELECTRODES

Therapy pads are not connected or are not
making good contact with the patient. Apply
therapy pads to the patient.

ECG FAULT: PACING IN FIXED MODE

The pacer is pacing in Fixed mode because
the ECG lead displayed in the primary
waveform is faulted.

PACING: SHORT DETECTED

The pacer output is short circuited due a test
plug connection or a device/MFC fault.