Setting upper and lower temperature alarm limits, Selecting the temperature label – ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 162

Advertising
background image

C

HAPTER

12 M

ONITORING

T

EMPERATURE

12-4

www.zoll.com

9650-001355-01 Rev. H

Setting Upper and Lower Temperature Alarm Limits

Initially, the Temperature Alarm Settings menu specifies that the Temperature alarms are

enabled (

ON

), and displays the default upper and lower limits. The upper and lower limits can

be

ON

or

OFF

(default is

OFF

). The following table lists the default Temperature limits for

adult, pediatric, and neonate patients, and gives the range in which you can set these limits:

Selecting the Temperature Label

Through the Temp Parameter Control Panel, you can configure alarms (as described
previously) and select descriptive Temperature channel labels that will appear on the numeric
display and on the trend report:

Figure 12-2 Temp Parameter Control Panel

You can select one of the following labels for each temperature channel:

Patient Type

Temperature Limit Default

Temperature Limit Range

Adult

Lower: 0.0

F (0.0° C)

Upper: 5.0

F (-15° C)

Lower: 0.0 - 89.8

 F (0.0 - 32.1° C)

Upper: 0.2 - 90.0

 F (-17.7 - 32.2° C)

Pediatric

Lower: 0.0

F (0.0° C)

Upper: 5.0

F (-15° C)

Lower: 0.0 - 89.8

 F (0.0 - 32.1° C)

Upper: 0.2 - 90.0

 F (-17.7 - 32.2° C)

Neonate

Lower: 0.0

F (0.0° C)

Upper: 5.0

F (-15° C)

Lower: 0.0 - 89.8

 F (0.0 - 32.1° C)

Upper: 0.2 - 90.0

 F (-17.7 - 32.2° C)

Temp

Lower

Upper

90

100

T1

T2

85

98

5

8

T1 Source Label

T2 Source Label

ǻ7$ODUP

T2 Alarm

T1 Alarm