Cleaning instructions

Cleaning the X Series unit

To clean the X Series unit, use a nearly dry cloth containing one of the mild cleaning agents
listed below. DO NOT allow cleaning agent or water to run into the crevices or connector
openings at any time. Thoroughly wipe off any excess cleaning solution from the X Series unit
with a dry cloth.

Always check monitor and connector opening for unusual wear, damage or dampness while
cleaning.

Use only these recommended cleaning agents:

Warm water

Hydrogen peroxide solution

Liquid soap

Wex-cide

®

Windex

®

T.B.Q.

®

Never use these cleaning agents:

Butyl alcohol

Denatured ethanol

Freon

Mild chlorine bleach solution

Isopropyl alcohol

Trichloroethane, trichloroethylene

Acetone

Vesphene II

Enviroquat

Staphene

Misty

Glutaraldehyde

Cleaning the NIBP Blood Pressure Cuff

Clean the cuff with common hospital disinfectants, including Clorox

®

(1:10 solution),

isopropyl alcohol, Lysol

®

solution, Phisohex

®

, Quatricide

®

, Virex

®

and Vesphene

®

. Wash

gently with the solution, then rinse. DO NOT allow solution to enter cuff tubes. Allow the cuff
and hose to completely dry before patient use.