AED Operation

9650-001355-01 Rev. H

X Series Operator’s Guide

13–7

3 Press SHOCK

WARNING!

Warn all persons in attendance of the patient to STAND CLEAR prior to defibrillator
discharge.

Do not touch the bed, patient, or any equipment connected to the patient during
defibrillation. A severe shock can result. Do not allow exposed portions of the patient’s
body to come in contact with metal objects, such as a bed frame, as unwanted pathways
for defibrillation current may result.

Press and hold the illuminated SHOCK button on the front panel until energy is delivered to
the patient.

Observe the patient or ECG response to verify that the shock has been delivered.

The delivered energy level and the shock number (1) are displayed in the panel at the bottom of
the screen.

Perform CPR

Begin chest compressions and rescue breathing per local protocol as prompted by the unit.

Note:

If ZOLL OneStep CPR pads, OneStep Complete pads or CPR-D-padz are connected,
the unit monitors the rate and depth of chest compressions and can display PUSH
HARDER
and GOOD COMPRESSIONS messages and voice prompts.

Repeat Analysis

After performing CPR for the configured period of CPR, the unit will restart ECG analysis
automatically.

Note:

Reanalysis of the ECG rhythm is inhibited for 3 seconds after a shock.

Stop CPR

After performing CPR for the configured period of CPR, the unit will issue a STOP CPR
prompt while it restarts ECG analysis.