Using the spo, Parameter control panel, Selecting the spco and spmet monitoring – ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 144

Advertising
background image

C

HAPTER

10 P

ULSE

CO-O

XIMETRY

(S

P

O2, SPCO,

AND

S

P

M

ET

)

10-10

www.zoll.com

9650-001355-01 Rev. H

Setting Upper and Lower SpCO and SpMet Alarm Limits

The following table lists the default SpCO and SpMet limits for adult, pediatric, and neonate
patients, and gives the range in which you can set these limits.

Using the SpO

2

Parameter Control Panel

Use the navigation keys to highlight and select the SpO

2

numeric display to display the

SpO2 Parameter Control Panel, on which you can select options that optimize the use of SpO

2

for the patient:

Figure 10-4 SpO2 Parameter Control Menu

Selecting the SpCO and SpMet Monitoring

If SpCO and SpMet are installed, the X Series unit allows you enable or disable SpCO and
SpMet monitoring.

Patient Type

SpCO Limit
Default

SpCO Limit Range

SpMet Limit
Default

SpMet Limit Range

Adult

Lower: 0%

Upper: 10%

Lower: 0- 98%

Upper: 2 - 100%

Lower: 0%

Upper: 3%

Lower: 0- 98%

Upper: 2 - 100%

Pediatric

Lower: 0%

Upper: 10%

Lower: 0 - 98%

Upper: 2 - 100%

Lower: 0%

Upper: 3%

Lower: 0- 98%

Upper: 2 - 100%

Neonate

Lower: 0%

Upper: 10%

Lower: 0 - 98%

Upper: 2 - 100%

Lower: 0%

Upper: 3%

Lower: 0- 98%

Upper: 2 - 100%

SpO2

Lower

Upper

SpO2 Alarm

85

100

SpCO Alarm

SpMet Alarm

On

On

8 seconds

Normal

Off

SpCO Monitoring

SpMet Monitoring

Averaging Time

Sensitivity

HR/PR Tone

0.0

0

10

3.0

Advertising