background image

3-21

IM 04L21B01-17EN

Commands

1

2

3

4

5

6

App

Index

TG

Sets a color scale band

Syntax

TG p1,p2,p3,p4,p5,p6,p7<terminator>
p1 Measurement or computation channel

number

p2 Area (OFF, IN, OUT)
p3 Display color (24 colors; see the SL

command, which is used to set a trip line)

p4 Lower display position limit
p5 Upper display position limit
p6 Lower display position limit (exponent)
p7 Upper display position limit (exponent)

Query

TG[p1]?

Example Set the channel 005 color scale band to the

range from –1.0000 to 0.5000 V (2-V range), and
set the display color to green.
TG005,IN,GREEN,-10000,5000

Description • Parameters p6 and p7 are valid on models

with the Log scale (/LG1) option.

• Parameters p6 and p7 are valid when p1 is set

to a measurement channel.

• Configure the settings as shown below.

When using a channel that is not set to Log

scale (/LG1 option):
• Parameters p4 and p5 are the

corresponding channel’s span range (the
scaling range when scaling is enabled).

• Set the parameters so that p4 is less than

p5.

• Parameters p6 and p7 are Don’t care

values in setting commands. In queries,
these parameters will always be returned
as 0.

When using a channel that is set to Log scale

(/LG1 option):
• Parameters p4 and p5 are the mantissas of

the upper and lower limits, respectively, of
the color scale band.

• Parameters p6 and p7 are the exponents of

the upper and lower limits, respectively, of
the color scale band.

• The range defined by the display position

lower limit and the display position upper
limit is 1.00E–16 to 1.00E+16.

Mantissas p4 and p5
100 to 999: The corresponding values are

1.00 to 9.99.

Exponents p6 and p7
Parameter p6 must be between –16 and

16 when p4 is 100. Parameter p6 must be
between –16 and 15 when p4 is a value
other than 100. The same conditions hold
for the relationship between parameters p7
and p5.

• Set the parameters so that the lower

display position limits (p4 and p6) are less
than their upper display position limits (p5

and p7).

SQ

Sets the LCD brightness and the

screen backlight saver

Syntax

SQ p1,p2,p3,p4<terminator>
p1 LCD brightness

1 to 8

p2 Screen backlight saver type

OFF

Disables the saver function

DIMMER Dims the backlight
TIMEOFF Turns off the backlight

p3 Amount of time until the FX switches to

saver mode
1MIN, 2MIN, 5MIN, 10MIN, 30MIN, 1H

p4 Event that causes the FX to return from

saver mode
KEY

Pressing of a key

KEY+ALM Pressing of a key or occurrence

of an alarm

Query

SQ?

Example Set the LCD brightness to 2 and the screen

backlight saver type to DIMMER. Set the amount
time of until the FX switches to saver mode to
5 minutes and the event that causes the FX to
return from saver mode to the pressing of a key.
SQ2,DIMMER,5MIN,KEY

Description If p2 is set to OFF, do not set p3 or p4.

TC

Sets the background color

Syntax

TC p1,p2<terminator>
p1 Screen (WHITE, BLACK)
p2 Historical trend screen (WHITE, CREAM,

LIGHTGRAY, BLACK)

Query

TC?

Example Set the screen background color to black and

the historical trend screen background color to
cream.
TCBLACK,CREAM

TP

Sets automatic display group

switching

Syntax

TP p1<terminator>
p1 Auto switching interval (5S, 10S, 20S, 30S,

1MIN)

Query

TP?

Example Switch between display groups at 5-second

intervals.
TP5S

3.4 Setting Commands