Measured and computed data – Yokogawa Value Series FX1000 User Manual

Page 144

Advertising
background image

4-8

IM 04L21B01-17EN

Measured and Computed Data

• The FD command is used to output the data.
• You can use the CB command to specify whether to output the data of measurement

channels set to skip and computation channels set to OFF.

Syntax

The measured/computed data is output in the following syntax along with the date and

time information for each channel.
EACRLF
DATE_yy/mo/ddCRLF
TIME_hh:mm:ss.mmmtCRLF
s_ccca

1

a

2

a

3

a

4

uuuuuufdddddE-ppCRLF

····························
ENCRLF

yy Year (00 to 99)
mo Month (01 to 12)
dd Day (01 to 31)
hh Hour (00 to 23)
mm Minute (00 to 59)
ss Second (00 to 59)
mmm Millisecond (000 to 999. A period is placed between seconds and

milliseconds.)

t Reserved (Space.)
s Data status (N, D, S, O, E, or B)

N: Normal
D: Differential input
S: Skip
O: Over
E: Error
B: Burnout

ccc Channel number (3 digits)

001 to 012: Measurement channel
101 to 124: Computation channel

a

1

a

2

a

3

a

4

a

1

Alarm status (level 1)

a

2

Alarm status (level 2)

a

3

Alarm status (level 3)

a

4

Alarm status (level 4)

(Each status is set to

H, L, h, l, R, r, T, t, or space.)

((

H: high limit alarm, L: low limit alarm, h: difference high-limit alarm, l:

difference low-limit alarm,

R: high limit on rate-of-change alarm, r: low

limit on rate-of-change alarm,

T: delay high limit alarm, t: delay low limit

alarm, space: no alarm)

uuuuuu Unit information (6 characters, left-justified)

mV____:

mV

V_____:

V

^C____:

°C

xxxxxx:

(User-defined character string)

f Sign (+, –)

4.2 Output Format of Text Data

Advertising