Lo loads setup data for setting mode, Li saves setup data – Yokogawa Value Series FX1000 User Manual

Page 114

Advertising
background image

3-32

IM 04L21B01-17EN

Example Switch the FX to basic setting mode.

DS1

Description • You cannot set p1 to 1 when the FX is

recording (memory sampling) or computing, is
formatting an external storage medium, or is
storing data to an external storage medium.

• You cannot set p1 to 1 when the FX is

formatting an external storage medium or is
storing data to an external storage medium.

• To apply settings that you have changed using

the basic setting commands, you need to save
the settings using the XE command. Be sure
to save the settings using the XE command
before you change the execution mode
from the basic setting mode to the operation
mode. Otherwise, the new settings will not be
applied.

LO

Loads setup data for setting

mode

Syntax

LO p1,p2<terminator>
p1 File name (up to 32 characters)
p2 Medium

0

CF slot

1

USB

Example Load the setting mode setup data from a setup

file named SETFILE1.
LOSETFILE1

Description • When you specify the file name, do not specify

the extension.

• You can set p2 to 1 on models that have the

USB interface (/USB1 option).

• If you omit parameter p2, the medium is set to

the CF slot.

• You cannot use this command to load basic

setting mode setup data. To load setting mode
setup data and basic setting mode setup data,
use the YO command.

• You cannot use this command when there is

no external storage medium inserted in the
FX.

• You cannot use this command on models that

do not have a CF card slot or USB interface
(/USB1 option).

LI

Saves setup data

Syntax

LI p1<terminator>
p1 File name (up to 32 characters)
p2 Medium

0

CF slot

1

USB

Example Save setting mode setup data and basic setting

mode setup data to a file named SETFILE2 on
the CF card.
LISETFILE2

Description • When you specify the file name, do not specify

the extension.

• You can set p2 to 1 on models that have the

USB interface (/USB1 option).

• If you omit parameter p2, the medium is set to

the CF slot.

• An extension is appended to the file name

when you save the file.

• You cannot use this command when there is

no external storage medium inserted in the
FX.

• You cannot use this command on models that

do not have a CF card slot or USB interface
(/USB1 option).

CM

Sets communication input data

(/M1, /PM1, and /PWR1 options)

Syntax

CM p1,p2<terminator>
p1 Communication input data number
p2 Communication input data

The ranges are –9.9999E+29 to
–1.0000E–30 and 1.0000E–30 to
9.9999E+29. You can also specify 0.

Five significant digits

Query

CM?

Example Set communication input data C01 to

1.0000E–10.
CMC01,1.0000E-10

EM

Starts or stops the e-mail

transmission function (/C7

option)

Syntax

EM p1<terminator>
p1 Action type

0

Start

1

Stop

Example Start the e-mail transmission function.

EM0

Description To use the e-mail transmission function, you

must configure the Ethernet interface, set e-mail
addresses, and enter the contents you want to
transmit.

3.5 Control Commands

Advertising