4-24

IM 04L21B01-17EN

Status Information

• The IS command is used to output the data. The output format varies between IS0

and IS1.

• The operation status of the recorder is output.
• For details on the status information, see section 5.2, “Bit Structure of the Status

Information.”

Output for the IS0 command

• Syntax

EACRLF
aaa.bbb.ccc.dddCRLF
ENCRLF

aaa Status information 1 (000 to 255)
bbb Status information 2 (000 to 255)
ccc Status information 3 (000 to 255)
ddd Status information 4 (000 to 255)

• Example

EA
000.000.032.000
EN

Output for the IS1 Command

• Syntax

EACRLF
aaa.bbb.ccc.ddd.eee.fff.ggg.hhhCRLF
ENCRLF

aaa Status information 1 (000 to 255)
bbb Status information 2 (000 to 255)
ccc Status information 3 (000 to 255)
ddd Status information 4 (000 to 255)
eee Status information 5 (000 to 255)
fff Status information 6 (000 to 255)
ggg Status information 7 (000 to 255)
hhh Status information 8 (000 to 255)

• Example

EA
000.000.032.000.000.000.000.000
EN

• Status information 3, 4, 7, and 8 are edge operation. They are cleared when read by

the IS command.

• Status information 1, 2, 5, and 6 are level operation. They are not cleared when read.

They are cleared when the event clears.

• The status information is made up of bits that correspond to each event. Each bit can

be turned ON/OFF with a filter.

• If an event occurs for a bit set to OFF by the filter, status information 3, 4, 7, and 8

discard the event. Status information 1, 2, 5, and 6 hold the event.

• The default filter setting is all ON.

4.2 Output Format of Text Data

Advertising