4-7

IM 04L21B01-17EN

Responses

1

2

3

4

5

6

App

Index

• Example

EA
SR001,VOLT,20mV,0,20
SR002,VOLT,20mV,0,20
···················
EN

Decimal Place/Unit Information

• The FE command is used to output the data.
• You can use the CB command to specify whether to output the data of measurement

channels set to skip and computation channels set to OFF.

•

Syntax

The data is output for each channel in the following syntax.

EACRLF
s_cccuuuuuu,ppCRLF
··················
ENCRLF

s Data status (N, D, or S)

N: Normal
D: Differential input
S: Skip (When the measurement range is set to SKIP for a

measurement channel or when the channel is turned OFF for a
computation channel)

ccc Channel number (3 digits)

001 to 012: Measurement channel
101 to 124: Computation channel

uuuuuu Unit information (6 characters, left-justified)

mV____:

mV

V_____:

V

^C____:

°C

xxxxxx:

(User-defined character string)

pp Decimal place (00 to 04)
No decimal (

00000) for 00.

One digit to the right of the decimal (

0000.0) for 01.

Two digits to the right of the decimal (

000.00) for 02.

Three digits to the right of the decimal (

00.000) for 03.

Four digits to the right of the decimal (

0.0000) for 04.

When you are using a channel that is set to Log scale (/LG1 option), the

decimal place of the FX digital display’s mantissa is displayed.