App index – Yokogawa Value Series FX1000 User Manual

Page 111

Advertising
background image

3-29

IM 04L21B01-17EN

Commands

1

2

3

4

5

6

App

Index

Setting the function menu

p1

Type (FUNC)

p2 and additional parameters Menu item to display
The FX displays the functions that you select

from below in the specified order on the menu.

Menu items that are not specified are not

displayed.

ALARMACK

AlarmACK

MESSAGE

Message

FREE_MESSAGE

Free message

MEDIA_EJECT

Media eject

SNAPSHOT
MANUAL_SAMPLE

Manual sample

TRIGGER

Event trigger

SAVE_DISPLAY

Saves display data

SAVE_EVENT

Saves event data

SAVE_STOP

Interrupts a save
operation

MATH_START/STOP Starts or stops

computation

MATH_RESET

Resets computation

MATH_ACK

Computation
data dropout
acknowledgment

TIMER_RESET

Timer reset

KEYLOCK

Locks or unlocks the
keys

LOGOUT

Logout

PASSWORD_CHANGE Password change
RATE_CHANGE

Switches between
display rate 1 and
display rate 2

BATCH
TEXT_FIELD

Text field display

JUMP_DISPLAY

Registers the screen
to return to

SYSTEM_INFO

System information

NETWORK_INFO

Network information

SNTP
EMAIL_START/STOP Starts or stops

sending E-mail

EMAIL_TEST

E-mail transmission
test

FTP_TEST

Example Display FREE MESSAGE and SNAPSHOT on

the function menu in that order.
SMFUNC,FREE_MESSAGE,SNAPSHOT

Description • If you specify the same menu item multiple

times, a command error will occur.

• You cannot specify “SEPARATOR.”
• You cannot use delimiters to omit parameters

(, ,).

• You cannot hide the “LOGOUT” menu item. If

you do not include it in the parameters, it will
be displayed as the last item.

• You can specify SNAPSHOT and

MANUALSAMPLE regardless of the FX
specifications. However, depending on
the FX specifications and the settings, the
corresponding functions may not operate.

• You cannot specify MEDIA_EJECT and

SAVE_STOP on models that do not have a CF
card slot or USB interface (/USB1 option).

• You cannot specify NETWORK_INFO, SNTP,

EMAIL_START/STOP, EMAIL_TEST, and
FTP_TEST on models that do not have the
Ethernet interface (/C7 option).

• You cannot specify MATH_START/STOP,

MATH_RESET, or MATH_ACK on models that
do not have the /M1, /PM1, or /PWR1 math
option.

Query

SM?

When you want to query all menu items.

SMDISP_MAIN?

When you want to query all main menu
items.

SMDISP_SUB?

When you want to query all sub menu items.

SMDISP_SUB,TREND?

When you want to query the trend sub menu
items.

SMFUNC?

When you want to query all function menu
items.

3.4 Setting Commands

Advertising