Yv sets an e-mail recipient address (/c7 option), Yw sets the e-mail sender address (/c7 option), Yx sets the e-mail smtp server name (/c7 option) – Yokogawa Value Series FX1000 User Manual

Page 124

Advertising
background image

3-42

IM 04L21B01-17EN

When you want to send system notifications

Syntax

YU p1,p2,p3,p4,p5,p6,p7<terminator>
p1 Information to send (SYSTEM)
p2 Recipient 1 (ON, OFF)
p3 Recipient 2 (ON, OFF)
p4 Inclusion of the source URL (ON, OFF)
p5 Subject (up to 32 characters)
p6 Header 1(up to 64 characters)
p7 Header 2(up to 64 characters)

Query

YU[p1]?

Example Send system notification e-mails that include

the source URL to recipient 1. Set the subject to
“SystemAlert” and header 1 to “LP2.”
YUSYSTEM,ON,OFF,ON,SystemAlert,LP2

When you want to send report generation

notifications

Syntax

YU p1,p2,p3,p4,p5,p6,p7<terminator>
p1 Information to send (REPORT)
p2 Recipient 1 (ON, OFF)
p3 Recipient 2 (ON, OFF)
p4 Inclusion of the source URL (ON, OFF)
p5 Subject (up to 32 characters)
p6 Header 1(up to 64 characters)
p7 Header 2(up to 64 characters)

Query

YU[p1]?

Example Send report generation notification e-mails that

include the source URL to recipient 1. Set the
subject to “Report” and header 1 to “LP2.”
YUREPORT,ON,OFF,ON,Report,LP2

Description • For details about system notifications, see

section 1.4.

• You can use report generation notification on

models with the /M1, /PM1, or /PWR1 math
option.

• For details on e-mail settings, see section 1.4.

YV

Sets an e-mail recipient address

(/C7 option)

Syntax

YV p1,p2<terminator>
p1 Selects the recipient

1

Recipient 1

2

Recipient 2

p2 Recipient address (up to 150 characters)

Query

YV[p1]?

Example Set recipient 1 to “[email protected]” and

[email protected].”
YV1,[email protected] fxuser2@
fx1000.com

Description • To specify multiple recipients, separate each

recipient with a space.

• For details on e-mail settings, see section 1.4.

YW

Sets the e-mail sender address

(/C7 option)

Syntax

YW p1<terminator>
p1 Sender address (up to 64 characters)

Query

YW?

Example Set the sender address to “fxadv.”

YWfxadv

Description For details on e-mail settings, see section 1.4.

YX

Sets the e-mail SMTP server

name (/C7 option)

Syntax

YX p1,p2,p3<terminator>
p1 SMTP server name (up to 64 characters)
p2 Port number (0 to 65535)
p3 Authentication (OFF, POPBEFORESMTP,

AUTH)
OFF

Authentication is not
used

POPBEFORESMTP

POP before SMTP is
used

AUTH

SMTP authentication
is used

Query

YX?

Example Set the SMTP server to “smtp.recorder.com” and

the port to 25. Use POP3 authentication.
YXsmtp.recorder.com,25,POPBEFORESMTP

Description For details on e-mail settings, see section 1.4.

YJ

Sets the Modbus client’s

destination server (/C7 option)

Syntax

YJ p1,p2,p3,p4,p5<terminator>
p1 Server number (1 to 16)
p2 Port number (0 to 65535)
p3 Host name (up to 64 characters)
p4 Unit number registration

AUTO

Do not use the unit number

FIXED Use a fixed unit number

p5 Unit number (0 to 255)

Query

YJ[p1]?

Example For server number 3, set the port number to

502, the host name to “fx1000,” the unit number
registration to FIXED, and the unit number to
127.
YJ3,502,fx1000,FIXED,127

3.6 Basic Setting Commands

Advertising