Measured/computed data and fifo data – Yokogawa Value Series FX1000 User Manual

Page 166

Advertising
background image

4-30

IM 04L21B01-17EN

Measured/Computed Data and FIFO Data

• The FD command is used to output the measured/computed data.
• The FF command is used to output the FIFO data.
• You can use the CB command to specify whether to output the data of measurement

channels set to skip and computation channels set to OFF.

• The ID number of the output format is 1. See “ID” in section 4.1.
• When you are using a channel that is set to Log scale (/LG1 option), the logarithmic

values are output. You can convert logarithmic values to physical values.

• About logarithmic values
The relationship between the physical value (V) and the logarithmic value (V’) is

shown below.
V = 10^(V’/1000)
V’ = 1000 × logV

However, V’ must be greater than or equal to –30000 and less than or equal to 30000,

and V must be greater than or equal to 1.00E–30 and less than or equal to 1.00E+30.

The special data of V’ (such as Skip, +Over, and –Over) may exceed 30000 or be less

than –30000. This special data is output as is. For details on the special data, see the
“Special Data Values” table on the following page.
Examples:

Physical Value (V)

logV

Logarithmic Value (V’)

1.00E–30

–30

–30000

2.00E–02

–1.699

–1699

2.00E+05

5.301

5301

1.00E+30

30

30000

–OVER

0x8001

+OVER

0x7FFF

2 byte

Number of blocks

Block 1

Block n

Number of bytes

2 byte

Binary data

(The binary data section on the

“Conceptual diagram” in

section 4.1.)

Number of Blocks

This is the number of blocks.

Number of Bytes

This is the size of one block in bytes.

Block

Type

Type

Year

Month

Channel

Channel

Day

A2A1

A2A1

Hour

A4A3

A4A3

Min

ms

Summer/

winter

Flag

s

Measured data

Computed data

1 byte 1 byte 1 byte 1 byte 1 byte

1 byte

2 bytes

1 byte

1 byte

4 bits

12 bits

4.3 Output Format of Binary Data