Fc outputs screen image data, Fe outputs setup data – Yokogawa Value Series FX1000 User Manual

Page 130

Advertising
background image

3-48

IM 04L21B01-17EN

3.8 Output Commands

(Setting, measured, and

computed data output)

FC

Outputs screen image data

Syntax

FC p1<terminator>
p1 GET (output screen image data)

Example Output screen image data from the FX.

FCGET

Description The FX captures the currently displayed screen

and outputs the data in PNG format.

FE

Outputs setup data

Syntax

FE p1,p2,p3<terminator>
p1 Output data type

0

Setting mode setup data

1

Decimal place and unit information

2

Basic setting mode setup data

4

Setup data file

5

Setup channel information output

6

Setup alarm information output

p2 First channel number (measurement

channel or computation channel)

p3 Last channel number (measurement

channel or computation channel)

Example Output the setting mode setup data of channels

001 to 005 from the FX.
FE0,001,005

Description • Make sure that the last channel number is

greater than or equal to the first channel
number.

• Parameters p2 and p3 are valid when p1 is

set to 0, 1, 2, 5, or 6. If you omit p2 or p3, the
command will specify all channels.

• Set p2 and p3 by referring to the table in

section 3.3.

FD

Outputs the most recent

measured data and computed

data

Syntax

FD p1,p2,p3<terminator>
p1 Output data type

0

The most recent measured data
and computed data in text format

1

The most recent measured data
and computed data in binary
format

6

Relay status and internal switch
status

p2 First channel number (measurement

channel or computation channel)

p3 Last channel number (measurement

channel or computation channel)

Example Output the most recent measured data and

computed data for channels 001 to 005 from the
FX in text format.
FD0,001,005

Description • The most recent measured data and computed

data means the most recent measured data
and computed data in the internal memory
when the FX receives the FD command.

• Make sure that the last channel number is

greater than or equal to the first channel
number.

• Parameters p2 and p3 are valid when p1 is set

to 0 or 1. If you omit p2 or p3, the command
will specify all channels.

• Set p2 and p3 by referring to the table in

section 3.3.

3.8 Output Commands (Setting, measured, and computed data output)